11 indexes, 1 explanation of the messages, 1 error messages – YSI AmmoLyt Plus 700 IQ User Manual
Page 61: 11 indexes -1, 1 explanation of the messages -1, 1 error messages -1
AmmoLyt
®Plus
700 IQ
Indexes
11 - 1
ba76003e01
01/2012
11
Indexes
11.1
Explanation of the messages
This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S
ENSOR
N
ET
system for the AmmoLyt
®Plus
700 IQ sensor.
Note
Information on
the contents and structure of the log book and
the structure of the message code
is given in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system
operating manual.
Note
The last three digits of the message code identify the source of the
message:
522 = AmmoLyt
®Plus
700 IQ (armature / component class, ADA
adapters)
395 = AmmoLyt
®Plus
700 IQ (ammonium / kalium sensor)
11.1.1 Error messages
Message code
Message text
EA1395
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA2522
Sensor temperature too high!
* Check process and application
EA3522
Sensor temperature too low!
* Check process and application
EAN395
Potassium measurement: range exceeded or undercut
* Check process
EIA522
Incorrect equipment
* for correct electrode equipment see operating manual
ES1522
Component hardware defective
* Contact service