2 setting table of the ammolyt, Setting table of the ammolyt+ -10 – YSI AmmoLyt Plus 700 IQ User Manual
Page 22

Commissioning
AmmoLyt
®Plus
700 IQ
3 - 10
ba76003e01
01/2012
3.5.2
Setting table of the AmmoLyt+
Menu item
Selection/values
Explanations
Measuring mode
NH4-N
NH4
mV
Citation form of the mass concentration or
voltage of the electrode.
Measuring range
(Measuring mode: NH4-N)
AutoRange
0.1 ... 100.0 mg/l
1 ... 1000 mg/l
2 measuring ranges can be selected.
With AutoRange, the instrument
automatically switches to the suitable
measuring range.
Measuring range
(Measuring mode: NH4)
AutoRange
0.1 ... 129.0 mg/l
1 ... 1290 mg/l
2 measuring ranges can be selected.
With AutoRange, the instrument
automatically switches to the suitable
measuring range.
Measuring range
(Measuring mode: mV)
-2000 ... 2000 mV
Fixed range
AmmoLyt+K
hide
display
hide (standard setting):
The potassium electrode is not displayed
as an extra sensor in the measured value
display.
display:
If the sensor is equipped with the
potassium electrode, the electrode is
displayed as the extra sensor AmmoLyt+K
in the measured value display.
The relevant settings can be done in an
extra setting table (see section 3.5.3).
Note: If the maximum number of sensors
for the system is exceeded by activating
the sensor, the AmmoLyt+K cannot be
activated.
Calib. history K
(only with the setting
AmmoLyt+K: hide)
Do not download
Transmit to log book
Transmit to log book generates a log book
message with the calibration history of the
potassium electrode. When opening the
setting table again the setting is reset to
Do not download.