8 technical data, Technical data, In chapter 8 t – YSI CR 4200 Thermoreactor User Manual
Page 33: Echnical, Chapter 8 t, 8technical data

CR 4200
Technical Data
33
ba76121e01
09/2012
8
Technical Data
Reactor type
Dry temperature control device
with safety cover
Cell shafts
2 x 12 cell shafts for reaction
cells 16 ± 0.2 mm
Reaction time setting
20 min, 30 min, 60 min, 120 min
(via fixed programs)
8 freely adjustable programs:
0...180 min
Temperature setting
100 °C, 120 °C, 148 °C, 150 °C
via fixed programs
and 8 freely adjustable
programs:
Room temperature ...170 °C
Controlling accuracy
± 1 °C ± 1 Digit
Temperature stability
± 0.5 K
Overtemperature protection
190 °C ± 5 °C
Heating time (with empty
thermoblock) from 25 °C to
100 °C approx. 5 min
120 °C approx. 7 min
148 °C approx. 10 min
Temperature of the enclose at
an environmental temperature
of 25 °C
< 30 °C with a block temperature
of 148 °C
Output
unidirectional RS232 interface
with 2 banana sockets for:
external temperature probe
or
PC cable
Power supply
230 VAC 50 Hz ± 15 %
115 VAC 60 Hz ± 15 %
Power consumption: 560 W
Fuses 2 x 6.3 AT
Enclosure
PC ABS, recyclable, high
temperature resistant
Protective class
I according to DIN VDE 0700
part 1/11.90