6 settings, 1 editing a temperature program, Settings 5.6.1 – YSI CR 4200 Thermoreactor User Manual
Page 26: Editing a temperature program, See section 5.6.1), Section 5.6.1)

Operation
CR 4200
26
ba76121e01
09/2012
5.6
Settings
5.6.1
Editing a temperature program
The temperature programs 1 to 8 are installed permanently and cannot
be changed.
The temperature programs 9 to 16 and the temperature test program
can be edited according to individual requirements.
Note
For the temperature test program, a TFK CR external temperature
probe is required (available as an accessory, see chapter 9
A
).
1
Switch to the Standby mode.
2
Keep the g key depressed while you press e to switch to the
SETUP menu.
SETUP and, in the second line, an editable parameter are
displayed.
3
Using 8 2 select a temperature program no. 9 to 16 or the
temperature test program.
4
Using g edit the selected temperature program.
The two parameters, temperature and time, are displayed.
The editing is marked on the display by *.
The selected parameter is marked by an arrow < or >.
5
Using 8 2 change the parameter (e.g. temperature ).
6
Using 4 6 switch to the other parameter.
7
Using 8 2 change the parameter (e.g. reaction time ).
8
Using g confirm the changes.
The marking on the display (*) disappears.