Operating modes
The thermoreactor has three operating modes:
Standby
The display shows the model and version number of the
thermoreactor.
Using the g and e keys simultaneously takes you to the SETUP
menu. There you can:
– Edit the 8 temperature programs and the temperature test
program (setting the temperature and reaction time for the
programs 9 to 16 and T, see section 5.6.1)
– Set the display contrast (CONTRAST:0 to 9, see section 5.6.2)
– Activate a manual confirmation before the timer for the reaction
time is started (START TIMER:MAN. or AUTO, see section
5.6.3)
Program selection
After switching on with e the display shows the current
temperature values for each thermoblock. The second display line
shows the temperature programs and the temperature test program
to be selected if a reaction time of at least one minute has been set
(see section 5.6.1). The control lamp above he display for the
selected thermoblock lights up green.
Active temperature program
The display shows the current temperature values for each
thermoblock. The control lamp above the display lights up or flashes
red.