What to do if...

57

ba75368e04

01/2011

Obviously incor-

rect

measured values

Instrument does

not react to key-

stroke

Display

to

Probe symbol

flashes

Message

StoFull

Cause

Remedy

pH electrode:

– pH electrode unsuitable

– Use suitable electrode

– Temperature difference

between buffer and test
sample too high

– Adjust temperature of buffers

or test samples

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Operating condition

undefined or EMC load
unallowed

– Processor reset:

Press the and
keys at the same
time and release them again.
The software version is
displayed.

Cause

Remedy

– Time-out of the interface

– Check the instrument that is

connected

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– All memory locations are full

– Output data storage and clear

data storage