Operation
handylab pH 12
44
ba75368e04
01/2011
4.5
Transmitting data
You have the following possibilities of transmitting data:
z One of the following options:
– With the AutoStore function (page 36), measured values are
periodically saved internally (save interval Int 1) and output on the
interface.
– With the Data transmission interval function (Int 2), measured values
are periodically output on the interface (see below).
z With the Output data storage function (page 38), calibration data or saved
measured values are output on the interface.
z Via the analog recorder output (page 46), measured values are output as
voltage values.
z With a communication kit (accessory), data can be transmitted bidirection-
Note
If you connect a recorder (analog output), the output on the digital interface
is switched off.
4.5.1
Data transmission interval (Int 2)
The interval to the data transmission (Int 2) determines the chronological in-
terval between automatic data transmissions. After the selected interval ex-
pires, the current data record is transmitted to the interface.
Note
When the AutoStore function is active, the data transmission is performed ac-
cording to the setting of the save interval (Int 1). Set the save interval (Int 1)
to OFF to activate the Data transmission interval (Int 2).