handylab pH 11
Operation
29
ba75366e04
01/2011
4.3.3
ConCal
Single-point
calibration
Use any buffer solution for this rapid method.
The calibration will be the more exact the nearer the pH value of the buffer
solution is to that of the test sample.
Two-point
calibration
For this procedure, use two buffer solutions:
z pH 7.0 ± 0.5
z any other buffer solution
Note
Skip the steps 2 and 9 if you use a pH electrode with a temperature sensor.
Note
At this point, the ConCal calibration can be broken off with
1
Press the
function display ConCal appears. The sensor symbol displays
the evaluation of the last calibration (or no sensor symbol in the
delivery state or after the measurement parameter has been
reset).
2
If necessary, enter the temperature of the first buffer solution
with <
▲> <▼>.
3
Immerse the pH electrode in the first buffer solution (pH 7.0 ±
0.5 in two-point calibration).
4
Press the
The measured pH value appears on the display.
5
Set the measured value to the nominal pH value of the buffer
solution (at the current temperature) with <
▲> <▼>.
6
Press the
(mV) and the sensor symbol appear on the display.
7
Press the