2 measuring the ph value, Measuring the ph value – Xylem handylab pH 11 User Manual
Page 17

handylab pH 11
Operation
17
ba75366e04
01/2011
4.2.2
Measuring the pH value
Setting the
resolution
The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the
while pressing the
AutoRead
(drift control)
The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.
For identical measurement conditions, the following criteria apply:
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the pH electrode in the test sample.
3
Press the
pH value appears on the display.
mV/pH
mol/L
8
Auto Store
TempError
RCL
LoBat
AutoCal DIN
CalError
ARng AR
AutoCal TEC ConCal
Time
Baud
Day.Month No.
Ident
Year
1
U
%
°F
S
mg/L
% ppm
ISE
8
4
2
2
99
6
°C
TP
pH
Reproducibility
Response time
Better than 0.02
> 30 seconds
1
Call up the pH measuring mode with
▲> <▼> .
2
Activate the AutoRead function with
The current measured value is frozen (hold function).
3
Start AutoRead with
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.
4
If necessary, start the next AutoRead measurement with
5
To terminate the AutoRead function: Press the