handylab pH 11

Operation

17

ba75366e04

01/2011

4.2.2

Measuring the pH value

Setting the

resolution

The measuring instrument shows the pH value with a resolution of 0.01 or
0.001. (Default setting is 0.01). To change the resolution, press the key
while pressing the key.

AutoRead

(drift control)

The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable impact on the reproducibility of the
measured values.

For identical measurement conditions, the following criteria apply:

1

Perform the preparatory activities according to section 4.2.1.

2

Immerse the pH electrode in the test sample.

3

Press the key until pH appears on the status display. The
pH value appears on the display.

mV/pH

mol/L

8

Auto Store

TempError

RCL

LoBat

AutoCal DIN

CalError

ARng AR

AutoCal TEC ConCal

Time

Baud

Day.Month No.

Ident

Year

1

U

%

°F

S

mg/L

% ppm

ISE

8

4

2

2

99

6

°C

TP

pH

Reproducibility

Response time

Better than 0.02

> 30 seconds

1

Call up the pH measuring mode with and <

> <> .

2

Activate the AutoRead function with .
The current measured value is frozen (hold function).

3

Start AutoRead with .
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.

4

If necessary, start the next AutoRead measurement with
.

5

To terminate the AutoRead function: Press the key.