Operation

103

4.2.3 TDS (total dissolved solids)

To measure the TDS, proceed as follows:

1 Perform the preparatory activities according to sec-

tion 4.2.

2 Immerse the conductivity measuring cell into the test

sample.

3 When measuring with an integrated temperature

probe continue with step 4.
When measuring without temperature probe:

– Determine the temperature of the test sample using

a thermometer

– Press the <

ì

> key repeatedly until

appears in

the status line

– Enter the temperature using <

> <

>.

4 Press the <

ì

> key repeatedly until TDS appears in

the status line. The TDS value of the sample appears
on the display.

5 Set up the TDS factor (0.40 ... 1.00) using <

> <

>.

6 Wait for a stable measured value.