Operation

100

AutoRead AR

(Drift control)

The

AutoRead

function (drift control) checks the stability of

the measurement signal. The stability has a considerable
effect on the reproducibility of the measured value.

.

Note
The current AutoRead measurement (with acceptance of
the current value) can be terminated at any time by pressing
.

1 Call up the measuring mode by pressing <

ì

>.

2 Activate the AutoRead function by pressing

.
The current measured value is frozen (Hold function).

3 Start the AutoRead function by pressing

.
AR flashes on the display until a stable measured val-
ue is reached.
This measured value is transmitted to the interface.

4 If necessary, start the next AutoRead measurement

by pressing .

5 To terminate the AutoRead function: Press the

key.