1 options for data transmission, Options for data transmission – Xylem Lab 960 User Manual
Page 45
![background image](/manuals/310563/45/background.png)
Lab 960
Operation
45
ba75557e05
11/2009
4.7.1
Options for data transmission
Via the USB interface you can transmit data to a PC. Via the RS 232
interface, you can transmit data to a PC or an external printer.
The following table shows which data are transmitted to the interface in
which way:
4.7.2
Automatically downloading measurement datasets at
intervals
In order to automatically download to the interface measured values at
certain time intervals, set the download interval (Int.2).
Setting the download
interval
The default setting for the download interval (Int.2) is OFF.
To switch the function on, set an interval (5 s, 10 s, 30 s, 1 min, 5 min,
10 min, 15 min, 30 min, 60 min):
Data
Control
Operation / description
Current measured
values
manual
z With
z Simultaneously with every
automatic storage process
(Int.1) (see page 40).
automatic, at
intervals
z With
_
>. Then you
can set the transmission
interval (Int.2) (see
page 45).
z Simultaneously with every
automatic storage process
(Int.1) (see page 40).
automatic
z After each measurement
with stability control.
Stored measured
values
manual
z All datasets
(see page 46)
calibration records
manual
z Without display indication
(see page 46).
z During the display
indication with
_
>
(see page 43).
automatic
z On completion of a
calibration procedure.
1
Press
_
> to open the setting of the Int.2 interval.