Appendix: firmware update, Firmware update – Xylem Lab 850, up to serial number 09470314 User Manual

Page 59

background image

Lab 850

59

ba75553e03

01/2006

Appendix: Firmware update

General information

With the Update_Labxxx_MxxxP program and a PC you can update
the firmware of the Lab 850 to the newest version.

For the update you need:

! a free RS232 interface on your PC
! the RS232 cable, Z390.

Program installation

With the installation program,
"Install_Update_Labxxx_MxxxP_Vx_yy_English.exe", install the
firmware update program on your PC.

Program start

Start the "Update_Labxxx_MxxxP" program from the Windows start
menu.

You can change the language via the language menu.

Firmware update

Proceed as follows:

After switching the meter off and on you can check whether the meter
has taken over the new software version (see page 45).

1

Connect the Lab 850 to a serial interface (COM port) of the PC
with the aid of the interface cable Z390.

2

Make sure the Lab 850 is switched on.

3

To start the update process click the OK button.
The program automatically recognizes the used interface.

4

To go on, follow the instructions of the program.
During the programming process, a corresponding message
and a progress bar (in %) appear.
The programming process takes approx. two minutes.
A terminatory message is displayed after a successful
programming process. The firmware update is now completed.

5

Disconnect the meter from the PC.
The instrument is ready for operation.