Lab 850

Operation

25

ba75553e03

01/2006

Displaying the

calibration data

You can display the data of the last calibration (see page 30).

Calibration evaluation

After calibrating, the meter automatically evaluates the calibration. The
zero point and slope are evaluated separately. The worse evaluation of
both is taken into account.The evaluation appears on the display as the
CalClock and in the calibration record.

Preparatory activities

CalClock

Zero point
[mV]

Slope
[mV/pH]

-15 ... +15

-60.5 ... -58

-20 ... +20

-58 ... -57

-25 ... +25

-61 ... -60.5
or
-57 ... -56

-30 ... +30

-62 ... -61
or
-56 ... -50

Clean the electrode
according to the
sensor operating
manual

CalError

< -30 or
> 30

... -62 or
... -50

Eliminate the error
according to chapter
6 W

HAT

TO

DO

IF

...

(page 43)

1

Switch on the meter with .

2

Connect a pH electrode to the meter.

3

Keep the buffer solutions ready.

4

Adjust the temperature of the solutions and measure the
current temperature if the measurement is made without
temperature sensor.

5

Set the buffer set to be used for calibration as necessary.