4.6

Reset

4.6.1

Resetting calibration values

This function resets the calibration values to the default condition. All
other meter settings are retained.

Calibration values in the

default condition

Note
The measuring system is not calibrated after a reset. Before measuring
recalibrate the meter.

Resetting calibration

values

Zero point

pH 7.000 (0 mV)

Slope

100 % (-59.16 mV/pH)

1

Press

_

> to open the menu for the reset of the

calibration data.
Ini.C is displayed.

2

Press <

▲><▼> to display no or YES.

YES: Reset the calibration values.
no: Retain the calibration values.

3

Confirm with .
The menu is finished.
The meter switches to the measuring mode.