beautypg.com

Xylem Lab 850, up to serial number 09470314 User Manual

Page 45

background image

Lab 850

What to do if...

45

ba75553e03

01/2006

Obviously incorrect

measured values

Instrument does not

react to keystroke

You want to know which

software

version is in the

instrument

Cause

Remedy

pH electrode:

– pH electrode unsuitable

– Use a suitable electrode

– Temperature difference

between buffer and test sample
too high

– Adjust temperature of buffer

or sample solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press and hold the
key and switch the meter on

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

During the display test,
display the software version
with .