2 Quality Control Certification

Date:

Customer: Order No.:

Pass

Model

GPR-1535 GB PPM Oxygen Transmitter S/N ________________


Sensor


( ) GPR-12-100-M ppm Oxygen Sensor
( ) XLT-12-100-M ppm Oxygen Sensor S/N___________________

Accessories

Owner’s Manual

Configuration

( ) A-1161-L1 PCB Assembly, Main / Display Software Ver. __________
A-1182 PCB Assembly, Interconnection
( ) BARR-1001 Barrier, Intrinsic Safety (ATEX Intrinsic Safety Certified)

Range(s): 0-10 PPM, 0-100 PPM, 0-1000 PPM, 0-1%, 0-25% (air calibration only)

Power: 12-24V DC two wire loop power

Barometric pressure compensation

B-3170 Glove Box Housing Assembly

Test - Electronics

Default zero (without sensor)

Default span @ 600uA or 300uA

Analog signal output 4-20mA full scale

Test - Gas Phase

Calibrates with adequate span adjustment within 10-50% FS

Baseline drift on zero gas <

± 2% FS over 24 hour period

Noise level <

± 1.0% FS

Span adjustment within 10-50% FS

Final

Overall inspection for physical defects

Options

Notes


When operated in conjunction with the manufacturer’s recommended optional intrinsic safety barriers the GPR-1535 GB meets
the intrinsic safety standards required for use in Class 1, Division 1, Groups C, D hazardous areas.

The GPR-1535 GB is also available in a version, requiring optional intrinsic safety barriers, that has been certified to ATEX
Directive 94/9/EC, Ex II 1 G Ex ia IIB T4 Tamb -20⁰C to + 50⁰C