Adjusting optical power – Ocean Optics FOS-2x2-TTL User Manual
Page 11
1: Setup
000-10000-040-02-1209
3
Figure 1: Rear of FOS Unit
Figure 2: Front of FOS Unit
Adjusting Optical Power
The FOS is factory optimized on a fiber-core-diameter of 200µm. If other core diameters are used, you
can change the focal points. Refer to Figures 1 and 2 when adjusting the optical power.
►
Procedure
Follow the steps below to change your focal points:
1. Move the shutter switch (4 – Figure 2) into the OPEN position to open the shutter.
2. Connect a fiber optic spectrometer or an optical power meter to one side of the FOS.
3. Connect a light source to the other side of the FOS.
4. Loosen the locking screw (5 – Figure 1) with a hexagonal socket screw key (SW 2.0 mm).
5. Move the SMA connector (2 – Figure 2) to adjust the optics of the FOS.
6. Tighten the locking screw (5 – Figure 1) when properly adjusted.
RIGHT
LEFT
5
3
1
2
4
6
- Apex 785 Raman (1 page)
- STS-UV (2 pages)
- TR2 Engineering Note (4 pages)
- SAD500 Communications and Control (19 pages)
- Red Tide USB650 Install (26 pages)
- Fiber Optic Termination Kit (6 pages)
- Transmissive pH Probe (10 pages)
- Remora (42 pages)
- PlasCalc (59 pages)
- Correcting Device Driver Issues (8 pages)
- ecoVis Krypton Light Source (16 pages)
- LPC-500CM (28 pages)
- HPX-2000 (24 pages)
- ADC1000-USB (27 pages)
- Torus Operating Instructions (30 pages)
- ADC2000-PCI (13 pages)
- Sensors for Real-Time Analysis (2 pages)
- IDRaman reader (2 pages)
- DH-2000-CAL (30 pages)
- QE65 Pro (32 pages)
- Collimating Lenses (2 pages)
- D1000 (2 pages)
- IDRaman mini (2 pages)
- HR2000CG-UV-NIR (42 pages)
- Cool Red (1 page)
- HL-2000 (20 pages)
- XE-1 Xenon (6 pages)
- USB-ADP Serial Adapters (3 pages)
- CHEM2000 (36 pages)
- Deuterium-Halogen Calibration Light Source (19 pages)
- NeoFox Engineering Note (30 pages)
- QE Pro (2 pages)
- OOIBase32 (140 pages)
- AR-1 Argon (6 pages)
- pH Sensor Patches, Probes and Cuvettes (36 pages)
- SpecLine Offline Spectroscopy (60 pages)
- HL-2000-HP-232 (26 pages)
- External Triggering Options Instructions for Spectrometers with Firmware Version 3.0 and Above (16 pages)
- Breakout Box (10 pages)
- USB-ISS-UV_VIS (4 pages)
- ISS-UV_VIS (6 pages)
- OOIColor (14 pages)
- LS-1 Series (12 pages)
- Apex Install (24 pages)
- DH-2000 (34 pages)