ISS-2 Installation and Operation Instructions
4
134-00000-000-01-201010
3. Remove the four screws that secure the fan to the base of the ISS-2. Take particular care to
save the washers for each screw (typically two per screw), as you will need to replace them
when replacing the screws.
Note
Two of the four screws also hold the front two legs of the ISS-2 in place.
4. Pull the fan gently (along with the legs) away from the lamp to remove the fan.
5. Loosen the setscrew underneath the fan with the included Allen wrench. This setscrew holds
the bulb in place.
Note
You do not need to remove the setscrew – loosening it is sufficient.
6. Remove the setscrews above each of the back legs of the ISS-2 using an Allen wrench (not
included). These screws keep the two halves of the lamp together.
Specifications
Path length:
1 cm
Collimating lens:
BK 7 glass (~360 nm – 2 µm*), 5 mm diameter, f/2
Collimating lens termination:
SMA 905
Filter slot:
Accepts filters up to ¼” (6 mm) thick
Base material:
Aluminum
Spectral range:
360 nm – 2 µm*
Dimensions (LxWxH):
8.9 cm (3.5”) x 5.1 cm (2”) x 3.2 cm (1.25”)
Power input:
12 VDC/800 mA – 2.1 mm center positive
Power output:
6.5 watts
Bulb life:
900 hours
Bulb color temperature:
3100K
Output to bulb:
5 volts/1.3 amps
Output regulation:
0.2% voltage
Time to stabilized output:
~ 30 minutes
Bulb output:
7400 foot-candles (7.4 MSCP)
* Though the product can be used to 2 µm, you can configure it to only “see” to 1100 nm with the
spectrometer.