OOINLCorrect Loading Non-Linearity Correction Coefficients Instructions

10. Choose Spectrometer | Linearize from the menu or click Non Linearity Correct to start the

experiment. You will see spectra on the graph, starting from low light intensities increasing up
to saturation. When the experiment is finished you will see a Linearity report displayed on the
screen.

11. On the Report Prompts dialog box, select OK. If you see a “file ‘#12.txt’ was not found in

application’s directory. Would you create a new file?” dialog box, choose New. Then close
the Linearity Report dialog box by selecting File | Close.

12. From the main menu choose Spectrometer | USB2000 Data. The USB2000 Internal

Parameters dialog box appears.

13. Click the Read USB2000 Internal Parameters button.

14. Click the Get Calculated Non-Linearity Values button.

15.

Click the Update USB2000 button to write the Non-Linearity correction coefficients to the
USB2000’s EEPROM.