CUV-VIS, CUV-UV Cuvette Holders Installation Instructions

043-00000-UV-01-201010

3

Installing the Fiber Supports

Procedure

1. Snap the clamps around the fibers after the fibers are screwed into the light source and

spectrometer.

2. Lift the clamps until they support the fibers.

3. Unsnap the clamps to remove the fibers.

Adjusting the Fit of the Cuvette

The CUV-ALL is preset for use with a 1-cm quartz cuvette. When properly adjusted, the cuvette
should fit snugly in the holder.

Procedure

1. Locate the two ball plunger screws. They are under two of the four collimating lenses.

2. Remove the fiber clamps for easier access to the screws.

3. Use the provided screwdriver to loosen the two ball plunger screws until the ball end of the

screws is just visible in the cell holder.

4. Insert your cuvette into the holder.

5. Tighten the ball plunger screws until the ball contacts the cuvette and starts to compress. Do

not over tighten.

Installing Filters

Procedure

1. Loosen the filter clamping screw with the provided screwdriver.

2. Insert the filter into the filter slot. The filter slot can accommodate filters up to 6 mm thick.

3. Clamp the filter in place by gently tightening the clamping screw finger tight.

Using the Temperature Stabilization Feature

This feature is used to heat or cool the cuvette holder base and cuvette.

Procedure

1. Remove the two plugs from the top side of the base. (The plug on the right side of the base

should stay in place but might require thread tape.)

2. Replace the plugs with the two barbed fittings (or any 1/8” NPT pipe thread fittings). Thread

tape might be required on the fittings.

3. Connect the fittings to a water source. Water will circulate through the base.