4

Acquisition

This worksheet page has buttons for the control of data acquisition, and displays the color space

values for the sample being analyzed.

Store Reference

In reflective color measurements, this button stores the reflection spectrum of either the white

standard (such as Ocean Optics’ WS-1) or your working standard or field tile (see the Parameters
page). For emissive color measurements, this button stores the spectrum of a blackbody of known

color temperature (e.g., Ocean Optics’ LS-1 has a color temperature of 3100K).

Store Dark

This button is for storing the spectrum obtained when there is no light entering the

spectrometer.

Calibrate Field Tile

This button is used to calibrate a working standard or field tile so that you can use something

other than a white standard to reference your spectrometer system.

Start Acquisition

After storing a dark and a reference, this button start the real-time data acquisition, which

constantly updates the color space values and graphs.

Stop Acquisition

This button stops the real-time data acquisition.

Log Value

This button takes the current XYZ and L*a*b* values and stores them in the Stored Values

worksheet page along with a time stamp and a brief, user-entered, description of the sample

being analyzed.