beautypg.com

Red tide specifications, Ccd detector specifications, Red tide spectrometer – Ocean Optics Red Tide USB650 Install User Manual

Page 21

background image

A: Specifications

170-00000-RT-02-1109

15

Item

Name

Description

9

OFLV
Filters

OFLV Variable Longpass Order-sorting Filters block second- and third-order light.
These filters are optional.

10

UV4
Detector
Upgrade

The detector’s standard window is replaced with a quartz window to enhance
spectrometer performance (<340 nm). This upgrade is optional.

Red Tide Specifications

The following sections provide specification information for the CCD detector in the Red Tide, as well as
the Red Tide Spectrometer itself.

CCD Detector Specifications

Specification

Value

Detector

Sony ILX511 CCD

No. of elements

2048 pixels

Sensitivity

75 photons per count at 400 nm

Pixel size

14 µm x 200 µm

Pixel well depth

~62,500 electrons

Signal-to-noise ratio

250:1 (at full signal)

A/D resolution

12 bit

Dark noise

3.2 RMS counts

Corrected linearity

>99.8%

Red Tide Spectrometer

Specification

Value

Dimensions

89.1 mm x 63.3 mm x 34.4 mm

Weight

190 g

Power consumption

90 mA @ 5 VDC