Resistance measurement – Prokit's Industries MT-2017 User Manual

Page 9

background image

8

The Maximum terminal voltage drop is 250mV except for the
10A range
.

Note: If
connected incorrectly with the voltage at these
ranges, quickly remove the test leads from the circuit as to
avoid damage to this tester.
(This tester can afford the voltage <250V DC/AC rms. for the
period of 5 seconds max.)

Resistance Measurement

WARNING: DO NOT APPLY VOLTAGE TO MEASURING
TERMINAL WHILE RANGE SWITCH IS IN OHM POSITION
.
1. Set the selector switch to the appropriate Ω range to be used.
2. Connect the BLACK test lead to the “-COM” jack and the RED

test lead to the Red “+” jack.

3. Short the leads by touching the probes together. Pointer should

read zero at the right hand end of the upper most scale, if it
doesn’t, use the Ohm adjust knob on the right hand of the panel
to line up the pointer with zero. (If pointer can’t be brought to
zero, replace battery.)

4. Connect the test leads across the resistance to be measured.
5. Take reading on the top “Ω” scale and multiply it by the

multiplication factor indicated by the selector switch.

6. If there is little or no pointer movement from the left side of the

scale, reset the selector switch to higher range. The effective
reading scope on an Ohm meter scale is within the area of
between 25 degree of Arc left side to the Midscale and 25
degree right side to the Midscale.

Note: If connected incorrectly with the voltage, quickly
remove the test leads from the circuit as to avoid damage to
this tester. (This tester can afford the voltage <250V DC/AC
rms. for the period of 5 seconds max.)