beautypg.com

Remarks – Prokit's Industries MT-2017 User Manual

Page 3

background image

2

Test

Functions

Range Accuracy

Remarks

DC V

0-0.1-2.5-10-50-
250 V-1000V

 3% FSD.

 4% FSD.
For 1000V

Input Impendence:
20KΩ/V Overload
Protection: Max.
1000V AC/DC
BUT 0.1V/2.5V/10V
250V Max.

Null DCV

±-5V, ±25V

 5% FSD.

Input Impedance: 40KΩ/V
Overload 1000V Max.

AC V

0-10-50-250V
-1000V

 4% FSD.

 5% FSD.
For 1000V

Input Impendence:
9KΩ/V Overload
Protection: Max.
1000V AC/DC
But 10V/50V only
250V Max. Band
width: 40 ~10K Hz

DC mA

0-0.05-2.5-25-25
0 mA, 10A

 3% FSD.

 4% FSD.
For 10A

Drop Voltage: 250 mV
Overload protected by
Fuses 0.5A/250V &
10A / 250V at 10A
range, and Oxide
Varactor.<250V AC/
DC(5s). Max. test time
1min. for 10A.

X 1: 0.2 ~ 2KΩ
Midscale at 20Ω
X 10: 2 ~ 20KΩ
Midscale at
200Ω
X 100: 20~
200KΩ Midscale
at 2000Ω
X1K: 200~ 2MΩ
Midscale at 0KΩ
X10K: 2K
~20MΩ Midscale
at 200KΩ

 4% of
ARC of
Scale
Length

Overload protected by
the Oxide Varactor &
Fuse <250V AC/DC
(5s).