Troubleshooting – causes and corrective measures, Icon function – Prokit's Industries NT-6570 User Manual
Page 9

TROUBLESHOOTING
– Causes and Corrective Measures
All message codes are displayed with either a numeric code or the
word "Error". The following errors can be corrected:
Code Cause
Corrective Measure
204
Calculation error
Refer to user menu, repeat the procedures
Temperature too high
Cool down instrument, use it in a stable
temperature environment
208
* Please operate this unit with outside temperature from 0°C to +40°C
220
Battery too low
Replace new batteries
253
Temperature too low
Warm up instrument
255 Receiver
signal
too
weak, measurement
time too long
Use target plate or change to a surface with
better reflectivity
256 Received
signal
too
strong
Target too reflective. Use target plate or a
surface that is less reflective
Error
Hardware error and
uncertainty error
Switch on/off the device several times. If the
symbol still appears, then your instrument is
defective. Please call your dealer for
assistance