beautypg.com

Diode measurement – Prokit's Industries MT-2018 User Manual

Page 9

background image

8

3. Short the leads by touching the probes together. Pointer should

read zero at the right hand end of the uppermost scale, if it
doesn’t, use the Ohm adjust knob on the right hand of the panel
to line up the pointer with zero. ( If pointer can’t be brought to
zero, replace battery.)

4. Connect the test leads across the resistance to be measured.
5. Take reading on the top “Ω” scale and multiply it by the

multiplication factor indicated by the selector switch.

6. If there is little or no pointer movement from the left side of the

scale, reset the selector switch to higher range. The effective
reading scope on an Ohm meter scale is within the area of
between 25 degree of Arc left side to the Midscale and 25
degree right side to the Midscale.


Note: If connected incorrectly with the voltage, quickly
remove the test leads from the circuit as to avoid the damage
to this tester.
(This tester can afford the voltage <250V DC/AC
rms. for the period of 5 seconds max.)

Diode Measurement

Set the selector switch to the appropriate Ω range to be used.
NOTE: To test the diode while current below 0.060 mA at X
10K range; current below 0.15 mA at X 1K range; current
below 1.5 mA at X 100 range; current below 15 mA at X 10
range; current below 150 mA at X 1 range.

For IF (forward current) test

Put the BLACK test lead to the “-COM” jack and the RED test
lead to the Red “+” jack. And then connect the Black probe to
the Positive terminal of the Diode, the Red probe to the
Negative terminal of the Diode.