VEGA VEGAPULS 61 (≥ 2.0.0 - ≤ 3.8) 4 … 20 mA__HART four-wire User Manual
Page 51
M
in. distance from antenna end
50
mm (1.969 in)
4)
R
ecommended measuring range, encapsu-
lated antenna system
up to 10 m (32.81 ft)
R
ecommended measuring range plastic horn
antenna
up to 20 m (65.62 ft)
Reference conditions to measuring accuracy (similar to DIN EN 60770-1)
R
eference conditions according to DIN EN 61298-1
-
T
emperature
+
18 … +30 °C (+64 … +86 °F)
-
R
elative humidity
45 … 75 %
-
A
ir pressure
860 … 1060
mbar/86 … 106 kPa (12.5 … 15.4 psig)
O
ther reference conditions
-
R
eflector
ideal reflector, e.g. metal plate 2 x 2 m
-
F
alse reflections
L
argest false echo 20 dB smaller than the useful
echo
Characteristics and performance data
F
requency
K-
band (26 GHz technology)
M
eas. interval approx.
1
s
B
eam angle -3 dB
5)
-
E
ncapsulated antenna system
22°
-
P
lastic horn antenna
10°
S
tep response or adjustment time
6)
<
1
s (dependent on the parameter settings)
M
ax. level change
A
djustable up to 1 m/min. (dependent on the
parameter settings)
M
ax. emitted HF power of the antenna system
-
P
ulse peak power
<
2
mW
-
P
ulse duration
<
2
ns
-
A
verage power
<
5 µW
-
A
verage power with 1 m distance
<
200
nW/cm
2
M
ax. emitted HF power of the antenna system - Version with increased sensitivity
-
P
ulse peak power
<
10
mW
-
P
ulse duration
<
2
ns
-
A
verage power
<
25 µW
-
A
verage power with 1 m distance
<
1 µW/
cm
2
4)
I
n products with low dielectric value up to 50 cm (19.69 in).
5)
C
orresponds to the range with 50 % of the emitted power
6)
T
ime to output the correct level (with max. 10 % deviation) after a sudden
level change.
VEGAPULS
61 • 4 … 20
mA/HART four-wire
51
10 S
upplement
28439
-EN
-100426