beautypg.com

VEGA VEGAPULS 66 (≥ 2.0.0 - ≥ 4.0.0) HART and accumulator pack User Manual

Page 58

background image

58

11 Supplement

VEGAPULS 66 • HART and accumulator pack

40933-EN-121011

Gas phase

Temperature

Pressure

1 bar

(14.5 psig)

10 bar

(145 psig)

50 bar

(725 psig)

100 bar

(1450 psig)

200 bar

(2900 psig)

Air

20 °C/68 °F

0.00 %

0.22 %

1.2 %

2.4 %

4.9 %

200 °C/392 °F -0.01 %

0.13 %

0.74 %

1.5 %

3.0 %

400 °C/752 °F -0.02 %

0.08 %

0.52 %

1.1 %

2.1 %

Hydrogen

20 °C/68 °F

-0.01 %

0.10 %

0.61 %

1.2 %

2.5 %

200 °C/392 °F -0.02 %

0.05 %

0.37 %

0.76 %

1.6 %

400 °C/752 °F -0.02 %

0.03 %

0.25 %

0.53 %

1.1 %

Steam (satu-

rated steam)

100 °C/212 °F 0.26 %

-

-

-

-

180 °C/356 °F 0.17 %

2.1 %

-

-

-

264 °C/507 °F 0.12 %

1.44 %

9.2 %

-

-

366 °C/691 °F 0.07 %

1.01 %

5.7 %

13.2 %

76.0 %

Characteristics and performance data
Measuring frequency

C-band (6 GHz technology)

Measuring cycle time approx.

350 ms

Step response time

1)

≤ 3 s

Tracking speed of the measuring window

max.

1 m/min

Beam angle

2)

dependent on the antenna system

Ʋ without horn antenna

38°

Ʋ ø 75 mm (2.953 in)

38°

Ʋ ø 96 mm (3.78 in)

30°

Ʋ ø 146 mm (5.748 in)

20°

Ʋ ø 196 mm (7.717 in)

17°

Ʋ ø 242 mm (9.528 in)

14°

Emitted HF power (depending on the parameter adjustment)

3)

Ʋ Average spectral transmission power

density

-31 dBm/MHz EIRP

Ʋ Max. spectral transmission power

density

+24 dBm/50 MHz EIRP

Ʋ Specific absorption rate (SAR)

0.47 mW/kg

Ambient conditions
Ambient, storage and transport tempera-

ture

-40 … +80 °C (-40 … +176 °F)

Process conditions
For the process conditions, please also note the specifications on the type label. The lowest value

always applies.

1)

Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk

solids applications, until the output signal has taken for the first time 90 % of the final value (IEC 61298-2).

2)

Outside the specified beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)

3)

EIRP: Equivalent Isotropic Radiated Power