7 maintenance and fault rectification, 1 maintenance, 2 rectify faults – VEGA VEGATOR 112 User Manual
Page 17
17
7 Maintenance and fault rectification
VEGATOR 112 •
46106-EN-131217
7 Maintenance and fault rectification
7.1 Maintenance
If the instrument is used properly, no special maintenance is required
in normal operation.
7.2 Rectify faults
A maximum of reliability is ensured. Nevertheless, faults can occur
during operation. These may be caused by the following, e.g.:
•
Measured value from sensor not correct
•
Voltage supply
•
Interference on the cables
The first measure to be taken is to check the input and output signals.
The procedure is described as follows. In many cases the causes can
be determined this way and faults can be easily rectified.
Should these measures not be successful, please call in urgent cases
the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since
we offer this service world-wide, the support is only available in the
English language. The service is free of charge, only the standard
telephone costs will be charged.
Depending on the reason for the fault and the measures taken, the
steps described in chapter "Set up" may have to be carried out again.
Cause
Rectification
Sensor not connected
correctly
– In Ex systems, make sure that the Ex protection
is not influenced by the measuring instruments
used.
– Measure the current and voltage on the connec-
tion cable to the sensor
– Faults in the sensor that cause a current change
under 0.05 mA or over 6.8 mA lead to a fault
signal with signal conditioning instruments.
– In idle state, the terminal voltage (without sensor)
is at least 8.2 V
– The signal conditioning instrument has an inner
resistance of 1 kOhm. With this, the voltages for
the different currents can be calculated.
Failure reasons
Fault rectification
24 hour service hotline
Reaction after fault recti-
fication
Red fault LED lights