VEGA VEGABAR 83 HART and accumulator pack - Operating Instructions User Manual
Page 58

58
11 Supplement
VEGABAR 83 • HART and accumulator pack
45035-EN-131107
Ambient conditions
Version
Ambient temperature
Storage and transport temper-
ature
Standard version
-40 … +80 °C (-40 … +176 °F)
-60 … +80 °C (-76 … +176 °F)
Version IP 66/IP 68 (1 bar)
-20 … +80 °C (-4 … +176 °F)
-20 … +80 °C (-4 … +176 °F)
Version IP 68 (25 bar) with connec-
tion cable PUR
-20 … +80 °C (-4 … +176 °F)
-20 … +80 °C (-4 … +176 °F)
Version IP 68 (25 bar), connection
cable PE
-20 … +60 °C (-4 … +140 °F)
-20 … +60 °C (-4 … +140 °F)
Process conditions
Product temperature piezoresistive/strain gauge measuring cell
Seal
Standard
Extended temperature
range
Version for oxygen ap-
plications
Without sealing (with pro-
cess fitting according to
EN 837)
-40 … +105 °C (-
40 … +221 °F)
-
-40 … +60 °C (-
40 … +140 °F)
FKM (VP2/A)
-20 … +105 °C (-
4 … +221 °F)
-20 … +150 °C (-
4 … +302 °F)
-20 … +60 °C
(+4 … +140 °F)
EPDM(A+P 75,5/KW75F) -30 … +105 °C (-
22 … +221 °F)
-30 … +150 °C (-
22 … +302 °F)
-30 … +60 °C (-
22 … +140 °F)
FFKM (Perlast G75S)
-15 … +105 °C
(+5 … +221 °F
-15 … +150 °C
(+5 … +302 °F)
-15 … +60 °C
(+5 … +140 °F)
FEPM (Fluoraz SD890)
-5 … +105 °C
(+23 … +221 °F)
-
-5 … +60 °C
(+23 … +140 °F)
Product temperature METEC
®
measuring cell
Version
Temperature range
Standard
-12 … +150 °C (+10 … +284 °F)
High temperature
-12 … +180 °C (+10 … +356 °F)
High temperature, screen sheet
-12 … +200 °C (+10 … +392 °F)
Mechanical stress, depending on the instrument version
Vibration resistance
4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration
with resonance)
Shock resistance
100 g, 6 ms according to EN 60068-2-27 (mechanical
shock)
Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry
Ʋ Cable gland
M20 x 1.5
Ʋ Cable diameter (options)
5 … 9 mm (0.20 … 0.35 in)
6 … 12 mm (0.24 … 0.47 in)
10 … 14 mm (0.39 … 0.55 in)