4 Mounting

VEGABAR 83 • HART and accumulator pack

45035-EN-131107

1

2

3

6

6

6

6

6

5

4

Fig. 6: Position of the filter element - non-Ex, Ex-ia and Ex-d-ia version
1 Single chamber housing plastic, stainless steel (precision casting)

2 Single chamber housing Aluminium

3 Single chamber housing, stainless steel electropolished

4 Double chamber housing plastic

5 Double chamber housing Aluminium

6 Filter element

With the following instruments a blind plug is installed instead of the

filter element:

Instruments in protection IP 66/IP 68 (1 bar) - ventilation via capil-

laries in fix connected cable

Instruments with absolut pressure

4.3 Process pressure measurement

Keep the following in mind when setting up the measuring system:

Mount the instrument above the measuring point

Possible condensation can then drain off into the process line.

Measurement setup in

gases