beautypg.com

VEGA VEGABAR 83 HART and accumulator pack - Operating Instructions User Manual

Page 34

background image

34

6 Set up with the display and adjustment module

VEGABAR 83 • HART and accumulator pack

45035-EN-131107

By entering a 4-digit PIN, you protect the sensor data against unau-

thorized access and unintentional modification.
In this menu item, the PIN is displayed or edited as well as modified.

However, it is only available when the adjustment is released in the

menu "Setup/ Lock/release adjustment ".

In delivery status, the PIN is "0000".

In this menu item, you adjust the internal clock of the sensor. There is

no adjustment to summer/winter time.

With a reset, certain parameter adjustments carried out by the user

are reset.

The following reset functions are available:
Delivery status: Restoring the parameter settings at the time of

shipment from the factory incl. the order-specific settings. A user-pro-

grammable linearization curve as well as the measured value memory

will be deleted.
Basic settings: Resetting the parameter settings incl. special

parameters to the default values of the respective instrument. A pro-

grammed linearization curve as well as the measured value memory

is deleted.
The following table shows the default values of the instrument. De-

pending on the instrument version or application, all menu items may

not be available or some may be differently assigned:

Reset - Setup

Menu item

Parameter

Default value

Measurement

loop name

Sensor

Application

Application pro-

cess pressure
Application level

No reset

Slave for elec-

tronic differential

pressure

No reset

Additional settings - PIN

Additional adjustments -

Date Time

Additional adjustments

- Reset