9 Diagnosis, asset management and service
VEGABAR 81 • 4 … 20 mA
45025-EN-131010
9.7 Software update
The following components are required to update the sensor soft-
ware:
•
Sensor
•
Voltage supply
•
Interface adapter VEGACONNECT
•
PC with PACTware
•
Current sensor software as file
You can find the actual sensor software as well as detailed infor-
mation of the procedur
"Software".
You can find information about the installation in the download file.
Caution:
Instruments with approvals can be bound to certain software ver-
sions. Therefore make sure that the approval remains effective with a
software update.
You can find detailed information on
"Approvals".
9.8 How to proceed in case of repair
You can find a repair form as well as detailed information on how to
pr
cates".
By doing this you help us carry out the repair quickly and without hav-
ing to call back for needed information.
If a repair is necessary, please proceed as follows:
•
Print and fill out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Please contact the agency serving you to get the address for
the return shipment. You can find the agency on our home page