beautypg.com

VEGA VEGABAR 81 4 … 20 mA - Operating Instructions User Manual

Page 30

background image

30

6 Set up with the display and adjustment module

VEGABAR 81 • 4 … 20 mA

45025-EN-131010

Unit of measurement:

If the level should be adjusted in a height unit, the density of the me-

dium must also be entered later during the adjustment.
In addition, the temperature unit of the instrument is specified. The se-

lection determines the displayed unit in the menu items "Peak value,

measuring cell temperature", "Peak value, electronics temperature"

as well as "HART variables".
Temperature unit:

Enter the requested parameters via the appropriate keys, save your

settings with [OK] and jump to the next menu item with the [ESC] and

the [->] key.

Especially with chemical seal systems, the installation position of the

instrument can shift (offset) the measured value. The position cor-

rection compensates this offset. Hence the actual measured value is

taken over automatically. With relative pressure measuring cells also a

manual offset can be carried out.

If with the automatic position correction, the actual measured value

should be taken over as corrective value, then this value must not be

influeces by product covering or a static pressure.
With the manual position correction, the offset value can be deter-

mined by the user. Select for this purpose the function "Edit" and

enter the requested value.
Save your settings with [OK] and move with [ESC] and [->] to the

next menu item.
When the position correction was carried out, then the actual meas-

ured value is corrected to 0. The corrective value appears with inverse

signs as offset value in the display.
The position correction can be repeated as often as necessary. How-

ever, if the sum of the corrective values exceeds 20 % of the nominal

measuring range, then no position correction is possible.

Setup - Position correc-

tion