9 Diagnostics and service
WEIGHTRAC 31 • Foundation Fieldbus
42375-EN-131120
Code
Text mes-
sage
Cause
Rectification
F036
Faulty pro-
gram memory
– Error during software
update
– Repeat software update
– Exchanging the electronics
F037
RAM hard-
ware error
– Error in RAM
– Restart instrument
– Exchanging the electronics
F038
Slave signals
failure
– Connection cable to the
Slave instrument interrupted
– Instrument not defined as
Slave instrument
– Check the connection cable
to the Slave instrument
– Define instrument as Slave
F040
Hardware
error
– Sensor defective
– Restart instrument
– Exchanging the electronics
F041
Photomultipli-
er error
– Error in the measured value
recording
– Restart instrument
– Exchanging the electronics
F052
Faulty config-
uration
– Invalid parameter adjust-
ment
– Carry out a reset
F053
Adjustment
data input not
correct
– Adjustment of the input not
correct
– Correct adjustment of the
input
F066
Faulty adjust-
ment
– Adjustment not yet carried
out
– Error during adjustment or
when entering the lineariza-
tion table
– Carrying out adjustment
– Carry out linearization
F068
Count rate too
high
– Faulty instrument settings
– Carry out a reset
F072
Limit exceed-
ed
– Faulty instrument settings
– Carry out a reset
F080
System error
– Instrument error
– Restart instrument
– Call our service
F086
Communica-
tion error
– Error in the Fieldbus com-
munication
– Restart instrument
– Call our service
F114
Error real time
clock
– Discharge accumulator
– Readjust real time clock