12 supplement, 1 technical data – VEGA VEGAPULS 63 (≥ 2.0.0 - ≥ 4.0.0) Modbus and Levelmaster protocol User Manual
Page 57
57
12 Supplement
VEGAPULS 63 • Modbus and Levelmaster protocol
41364-EN-130228
12 Supplement
12.1 Technical data
General data
316L corresponds to 1.4404 or 1.4435
Materials, wetted parts
Ʋ Hygienic antenna encapsulation
PTFE, TFM-PTFE, PFA
Ʋ Surface roughness of the antenna
encapsulation
R
a
< 0.8 µm
Ʋ Additional process seal with certain
hygienic fittings
FKM, EPDM
Materials, non-wetted parts
Ʋ Process fitting
316L
Ʋ Plastic housing
plastic PBT (Polyester)
Ʋ Aluminium die-casting housing
Aluminium die-casting AlSi10Mg, powder-coated - basis:
Polyester
Ʋ Stainless steel housing
316L
Ʋ Seal between housing and housing
cover
NBR (stainless steel housing, precision casting), silicone
(aluminium/plastic housing; stainless steel housng,
electropolished)
Ʋ Inspection window in housing cover
(optional)
Polycarbonate
Ʋ Ground terminal
316L
Ohmic contact
Between ground terminal and process fitting
Process fittings
Ʋ Flanges
DIN from DN 25, ANSI from 1"
Ʋ Hygienic fittings
Clamp, bolting according to DIN 11851, hygienic fitting
with saddle flange according to DIN 11864-2-A, SMS
Weight (depending on housing, process
fitting and antenna)
approx. 3.5 … 15.5 kg (4.409 … 33.95 lbs)
Required torque of the flange screws
60 Nm (44.25 lbf ft)
Max. torque for NPT cable glands and Conduit tubes
Ʋ Plastic housing
10 Nm (7.376 lbf ft)
Ʋ Aluminium/Stainless steel housing
50 Nm (36.88 lbf ft)
Input variable
Measured variable
The measured quantity is the distance between process
fitting of the sensor and product surface. The reference
plane is the seal surface on the process fitting or the
lower side of the flange.