Maintenance: Due to external influences, the instrument function is
limited. The measurement is affected, but the measured value is still
valid. Plan in maintenance for the instrument because a failure is
expected in the near future (e.g. due to buildup).

This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.

The following table shows the codes and text messages of the status
message "Failure" and provides information on causes as well as
corrective measures.

Code
Text mes-
sage

Cause

Rectification

PA Dev-
Spec
Diagnosis

F

013

no measu-
red value
available

l Sensor does not de-

tect an echo during
operation

l Antenna system con-

taminated or defecti-
ve

l Check or correct in-

stallation and/or pa-
rameter adjustment

l Clean or exchange

process component
or antenna

Bit 0

F

017

Adjustment
span too
small

l Adjustment not within

specification

l Change adjustment

according to the limit
values (difference
between min. and
max. ≥ 10 mm)

Bit 1

F

025

Error in the
linearization
table

l Index markers are not

continuously rising,
for examle unlogical
value pairs

l Check linearization

table

l Delete table/Create

new

Bit 2

F

036

No operab-
le software

l Failed or interrupted

software update

l Repeat software up-

date

l Check electronics

version

l Exchange of the

electronics

l Send instrument for

repair

Bit 3

F

040

Error in the
electronics

l Hardware defect

l Exchange of the

electronics

l Send instrument for

repair

Bit 4

F

080

l General software er-

ror

l Separate operating

voltage briefly

Bit 5

Failure (failure)