11 Supplement
VEGAPULS 67 • HART - integrated accumulator
40795-EN-121011
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone Ex factory
CET
Measurement electronics temerature
Resolution
1 °C (1.8 °F)
Accuracy
±1 °C (1.8 °F)
External battery charger
Mains voltage
100 … 240 V AC
Output voltage
24 V DC
Output current (shortcircuit proof) max. 500 mA
Load current limitation
70 mA
DC plug (inside plus, outside minus)
2.1 mm
Integrated accumulator
Type
Lithium ions
Voltage
14.8 V
Accumulator capacity
4.7 Wh
Charging period from 0 % to 100 %
approx.
4 h
Operating time after 10 min. charging of
0 %
> 3 h
Operating time in mode 4 (sensor per-
manently on) with completely charged
accumulator
> 60 h
Temperature range
Ʋ Charge accumulator
0 … +45° C (+32 … +167 °F)
Ʋ Accumulator operation
-20 … +60° C (-4 … +140 °F)
Temperature derating accumulator capacity
Ʋ +25° C (+77 °F)
100 %
Ʋ -10° C (+14 °F)
50 %
Electrical protective measures
Protection, depending on housing version
Ʋ Plastic housing
IP 66/IP 67
Ʋ Aluminium housing, stainless steel
housing - investment casting, stain-
less steel housing - electro-polished
IP 66/IP 68 (0.2 bar)
5)
Overvoltage category
III
Protection class
III
5)
The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.