beautypg.com

VEGA VEGASOURCE 35 User Manual

Page 42

background image

42

7 Maintenance and fault rectification

VEGASOURCE 35 •

38132-EN-130415

The action protects the concerned persons until the responsible ra-

diation safety officer arrives which can prescribe additional measures.
The person engaged with the supervision of the source (i.e. the

person stipulated and authorized by the plant operator) is responsible

that this procedure is maintained.

Determine the dangerous area on site by measuring the local dose

rate in µSv/h

Largely cordon off the concerned area with a yellow marking tape

or with a rope and mark the area with the international aler icon for

radiation

In this case, the source container must be dismounted.
Direct the exit channel to a thick wall (e.g. of steel or lead) or mount

an blind flange in front of the exit channel.
Persons may only stand behind the source container - never in front

of the radiation exit channel (flange of VEGASOURCE 35).
The transport lug on the housing facilitates safe handling.

In this case, the source must be kept secure in another place or an

additional shielding must be provided.
The source must only be transported with pliers or a gripper and the

distance to the body must be kept as large as possible.
The time required for transport should be estimated and optimized in

advance by tests and training.

Pass all necessary information immediately on to the responsible

local and national authorities

After thorough investigation of the condition on site, the responsi-

ble radiation safety officer must agree along with the local author-

ity a suitable corrective measure for the existing problem

Note:

National regulations can prescribe deviating procedures and notifica-

tion requirements.

The source holder cannot

be brought into position

AUS (OFF)

The source is no longer

in the source container

Information to the re-

sponsible authority