9 Diagnosis, Asset Management and service

VEGAPULS 66 enamel • Profibus PA

36529-EN-121011

Code
Text mes-

sage

Cause

Rectification

PA DevS-

pec
Diagnosis

F260
Error in the

calibration

– Error in the calibra-

tion carried out in the

factory

– Error in the EEPROM

– Exchanging the elec-

tronics

– Send instrument for

repair

Bit 9

F261
Error in the

configura-

tion

– Error during setup

– False signal suppres-

sion faulty

– Error when carrying

out a reset

– Repeat setup

– Repeat reset

Bit 10

F264
Installa-

tion/Setup

error

– Adjustment not within

the vessel height/

measuring range

– Max. measuring range

of the instrument not

sufficient

– Check or correct

installation and/or

parameter adjustment

– Use an instrument

with bigger measuring

range

Bit 11

F265
Meas-

urement

function

disturbed

– Sensor no longer

carries out a measure-

ment

– Operating voltage

too low

– Check operating

voltage

– Carry out a reset

– Separate operating

voltage briefly

Bit 12

The following table shows the error codes and text messages in the

status message "Function check" and provides information on causes

as well as corrective measures.

Code
Text mes-

sage

Cause

Rectification

C700
Simulation ac-

tive

– A simulation is active

– Finish simulation

– Wait for the automatic end

after 60 mins.

The following table shows the error codes and text messages in the

status message "Out of specification" and provides information on

causes as well as corrective measures.

Code
Text mes-

sage

Cause

Rectification

S600
Unpermissi-

ble electronics

temperature

– Temperature of the elec-

tronics in the non-specified

section

– Check ambient temperature

– Isolate electronics

– Use instrument with higher

temperature range

S601
Overfilling

– Danger of vessel overfilling

– Make sure that there is no

further filling

– Check level in the vessel

Function check

Out of specification