beautypg.com

VEGA VEGABAR 54 Foundation Fieldbus - climate compensated User Manual

Page 50

background image

50

10 Supplement

VEGABAR 54 • Foundation Fieldbus - climate compensated

36791-EN-130321

Thermal change of the current output
Applies also to the analogue 4 … 20 mA current output and refers to the set span.
Thermal change, current output

< 0.05 %/10 K, max. < 0.15 %, each with -40 … +80 °C

(-40 … +176 °F)

0,15 %

-0,15 %

-40°C

-20°C

20°C

40°C

60°C

80°C

Fig. 24: Thermal change, current output

Long-term stability (according to DIN 16086 and IEC 60770-1)
Applies to digital HART interface as well as to analogue current output 4 … 20 mA. Specifications

refer to the set span. Turn down (TD) is the relation nominal measuring range/set span.

Nominal measuring range in bar/kPa

Nominal measuring

range in psig

Long-term drift of the zero signal

0 … 0.2 bar/0 … 20 kPa

0 … 3 psig

< (1 % x TD)/year

0 … 0.4 bar/0 … 40 kPa

0 … 6 psig

< (1 % x TD)/year

0 … 1 bar/0 … 100 kPa

0 … 15 psig

< (0.25 % x TD)/year

0 … 2.5 bar/0 … 250 kPa

0 … 35 psig

< (0.25 % x TD)/year

0 … 5 bar/0 … 500 kPa

0 … 70 psig

< (0.1 % x TD)/year

0 … 10 bar/0 … 1000 kPa

0 … 150 psig

< (0.1 % x TD)/year

0 … 25 bar/0 … 2500 kPa

0 … 350 psig

< (0.1 % x TD)/year

0 … 60 bar/0 … 6000 kPa

0 … 900 psig

< (0.1 % x TD)/year

-1 … 0 bar/-100 … 0 kPa

-15 … 0 psig

< (0.25 % x TD)/year

-1 … 1.5 bar/-100 … 150 kPa

-15 … 25 psig

< (0.25 % x TD)/year

-1 … 5 bar/-100 … 500 kPa

-15 … 70 psig

< (0.1 % x TD)/year

-1 … 10 bar/-100 … 1000 kPa

-15 … 150 psig

< (0.1 % x TD)/year

-1 … 25 bar/-100 … 2500 kPa

-15 … 350 psig

< (0.1 % x TD)/year

-1 … 60 bar/-100 … 6000 kPa

-15 … 900 psig

< (0.1 % x TD)/year

-0.1 … 0.1 bar/-10 … 10 kPa

-1.5 … 1.5 psig

< (1 % x TD)/year

-0.2 … 0.2 bar/-20 … 20 kPa

-3 … 3 psig

< (1 % x TD)/year

-0.5 … 0.5 bar/-50 … 50 kPa

-7 … 7 psig

< (0.25 % x TD)/year

Ambient conditions
Ambient, storage and transport temperature

Ʋ Standard version

-40 … +80 °C (-40 … +176 °F)