10 Supplement
VEGABAR 54 • Foundation Fieldbus - climate compensated
36791-EN-130321
Nominal range
Overload capacity, max.
pressure
Overload capacity, min. pres-
sure
-1 … +60 bar/-100 … +6000 kPa
+200 bar/+20000 kPa
-1 bar/-100 kPa
-0.1 … +0.1 bar/-10 … +10 kPa
+35 bar/+3500 kPa
-1 bar/-100 kPa
-0.2 … +0.2 bar/-20 … +20 kPa
+35 bar/+3500 kPa
-1 bar/-100 kPa
-0.5 … +0.5 bar/-50 … +50 kPa
+50 bar/+5000 kPa
-1 bar/-100 kPa
Nominal measuring ranges and overload capability in psig
The specifications are only an overview and refer to the measuring cell. Limitations due to the ma-
terial and version of the process fitting are possible. The specifications on the nameplate apply.
Nominal range
Overload capacity, max.
pressure
Overload capacity, min. pres-
sure
Gauge pressure
0 … +3 psig
+500 psig
-15 psig
0 … +6 psig
+700 psig
-15 psig
0 … +15 psig
+700 psig
-15 psig
0 … +35 psig
+725 psig
-15 psig
0 … +70 psig
+950 psig
-15 psig
0 … +150 psig
+1300 psig
-15 psig
0 … +350 psig
+1900 psig
-15 psig
0 … +900 psig
+2900 psig
-15 psig
-15 … 0 psig
+500 psig
-15 psig
-15 … +25 psig
+700 psig
-15 psig
-15 … +70 psig
+950 psig
-15 psig
-15 … +150 psig
+1300 psig
-15 psig
-15 … +350 psig
+1900 psig
-15 psig
-15 … +900 psig
+2900 psig
-15 psig
-1.5 … +1.5 psig
+500 psig
-15 psig
-3 … +3 psig
+500 psig
-12 psig
-7 … +7 psig
+700 psig
-15 psig
Reference conditions and actuating variables (according to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
Ʋ Temperature
+15 … +25 °C (+59 … +77 °F)
Ʋ Relative humidity
45 … 75 %
Ʋ Air pressure
860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Determination of characteristics
Limit point adjustment according to IEC 61298-2
Characterstic curve
Linear
Reference installation position
upright, diaphragm points downward
Influence of the installation position
< 0.2 mbar/20 Pa (0.003 psig)