beautypg.com

2 supplementary information foundation fieldbus – VEGA VEGAPULS SR 68 Foundation Fieldbus User Manual

Page 72

background image

72

11 Supplement

VEGAPULS SR 68 • Foundation Fieldbus

38297-EN-130624

Ʋ EEx-ia instrument - Power supply

FISCO model

9 … 17.5 V DC

Ʋ EEx-ia instrument - Power supply

ENTITY model

9 … 24 V DC

Ʋ EEx-d instrument

14 … 32 V DC

Power supply by/max. number of sensors

Ʋ Fieldbus

max. 32 (max. 10 with Ex)

Electrical protective measures
Protection, depending on housing version

Ʋ Plastic housing

IP 66/IP 67

Ʋ Aluminium housing, stainless steel

housing - investment casting, stain-

less steel housing - electro-polished

IP 66/IP 68 (0.2 bar)

4)

Ʋ Aluminium and stainless housing, in-

vestment casting (optionally available)

IP 66/IP 68 (1 bar)

Overvoltage category

III

Protection class

III

Approvals
Instruments with approvals can have different technical data depending on the version.
For that reason the associated approval documents of these instruments must be carefully noted.

They are part of the delivery or can be downloaded under www.vega.com and "VEGA Tools" as

well as under "Downloads" and "Approvals".

11.2 Supplementary information Foundation Fieldbus

The following table gives you an overview of the instrument versions and the corresponding device

descriptions, the electrical characteristics of the bus system as well as the applied function blocks.

Revisions Data

DD-Revision

Rev_01

CFF-File

010101.cff

Device Revision

0101.ffo
0101.sym

Cff-Revision

xx xx 01

Device software revision

> 4.4.0

ITK (Interoperability Test Kit) Number

5.0.2

4)

The prerequisites for maintaining the protection rating are a suitable cable as well as correct mounting.