4 Mounting

VEGAPULS SR 68 • Foundation Fieldbus

38297-EN-130624

4 Mounting

4.1 General instructions

On instruments with process fitting thread, the hexagon must be tight-

ened with a suitable screwdriver. Wrench size see chapter "Dimen-

sions".

Warning:

The housing must not be used to screw the instrument in! Applying

tightening force can damage internal parts of the housing.

Protect your instrument further through the following measures

against moisture penetration:

Use the recommended cable (see chapter "Connecting to power

supply")

Tighten the cable gland

Loop the connection cable downward in front of the cable gland

This applies particularly to:

Outdoor mounting

Installations in areas where high humidity is expected (e.g. through

cleaning processes)

Installations on cooled or heated vessels

Make sure that all parts of the instrument exposed to the process are

suitable for the existing process conditions.
These are mainly:

Active measuring component

Process fitting

Process seal

Process conditions are particularly:

Process pressure

Process temperature

Chemical properties of the medium

Abrasion and mechanical influences

You can find the specifications of the process conditions in chapter

"Technical data" as well as on the nameplate.

4.2 Mounting preparations - Horn antenna

The instrument is also available in versions with an antenna whose

diameter is larger than the process fitting (thread, flange). In such

cases the antenna must be disconnected from the process fitting

before mounting. Proceed as follows:
1. Loosen the hexagon socket screws (3) on the antenna socket

with an Allen wrench (size 3)

2. Remove the antenna (4)
Note:

The plastic cone may not be pulled out of the antenna socket.

Screwing in

Protection against mois-

ture

Suitability for the process

conditions