beautypg.com

VEGA PLICSMOBILE T61 User Manual

Page 28

background image

28

7 Setup

PLICSMOBILE T61 • External GSM/GPRS radio unit

37700-EN-130228

For this option, an e-mail account with name of the inbox and outbox

server (POP3/SMTP) as well as the user name/password are required

for authentication. These data are available from your e-mail provider.

You can find further information in the online help of the PLICSMO-

BILE DTM.
The measurement data can be sent either directly in the e-mail or in

an attached file. The formats TXT, CSV, HTML or XML are available.

Fig. 14: Input of the e-mail access data

For measured value transmission via SMS, no additional access data

are required because all needed information is already available via

the mobile phone contract.

For the configuration, only the URL or the IP address of the WEB-VV

server are required. When hosting is carried out through VEGA, the

server is contacted via the URL: "web-vv.vega.com". If WEB-VV is

hosted in your company, you will get the URL from your IT depart-

ment.

The setup of PLICSMOBILE and sensor is carried out via USB and

a PC with PACTware and corresponding DTM. With a SIM card with

activated CSD service, access via the radio link is also possible

(remote parameter adjustment). Keep in mind that this is carried out

via a dial-in connection, which bases invoicing on connection time.

Considerable connection costs can be incurred depending on the

duration of use.

E-mail transmission

SMS transmission

WEB-VV

Remote parameter ad-

justment (Dial-In)