beautypg.com

VEGA VEGAPULS SR 68 4 … 20 mA_HART two-wire User Manual

Page 73

background image

73

11 Supplement

VEGAPULS SR 68 • 4 … 20 mA/HART two-wire

38294-EN-130624

Seal

Antenna impedance

cone

Process temperature (measured on the process

fitting)

FFKM (Kalrez 6230)

PFFE

-15 … +130 °C (5 … +266 °F)

PEEK

-15 … +250 °C (5 … +482 °F)

Vessel pressure - horn antenna

Ʋ Antenna impedance cone PTFE

-1 … 40 bar/-100 … 4000 kPa (-14.5 … 580 psi)

Ʋ Antenna impedance cone PEEK

-1 … 100 bar/-100 … 10000 kPa (-14.5 … 1450 psi)

Vessel pressure with swivelling holder

-1 … 1 bar (-100 … 100 kPa/-14.5 … 14.5 psig)

Vessel pressure relating to the flange

nominal stage

see supplementary instructions manual "Flanges ac-

cording to DIN-EN-ASME-JIS"

Vibration resistance

Ʋ Horn antenna

4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration

with resonance)

Ʋ Parabolic antenna

1 g at 5 … 200 Hz according to EN 60068-2-6 (vibration

with resonance)

Shock resistance

Ʋ Horn antenna

100 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

Ʋ Parabolic antenna

25 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

Data on rinsing air connection
Pressure max.

6 bar (87.02 psig)

Air quantity without reflux valve, with

Ʋ 0.2 bar (2.9 psig)

2.0 m

3

/h

Ʋ 0.5 bar (7.25 psig)

3.5 m

3

/h

Ʋ 1 bar (14.5 psig)

4.5 m

3

/h

Ʋ 1.5 bar (21.76 psig)

5 m

3

/h

Ʋ 2 bar (29.0 psig)

5.5 m

3

/h

Thread

G⅛ A

Closure with

Ʋ Non-Ex

Dust protection cover of PE

Ʋ Ex

Threaded plug of 316Ti

Reflux valve - unmounted (as option with non-Ex version, included in the scope of delivery with Ex

version)

Ʋ Material

316Ti

Ʋ Seal

FKM (SHS FPM 70C3 GLT), FFKM (Kalrez 6375)

Ʋ for tube diameter

6 mm

Ʋ opening pressure

0.5 bar (7.25 psig)

Ʋ Nominal pressure stage

PN 250

Electromechanical data - version IP 66/IP 67 and IP 66/IP 68; 0.2 bar
Options of the cable entry