beautypg.com

VEGA VEGAPULS 63 (≥ 2.0.0 - ≥ 4.0.0) HART and accumulator pack User Manual

Page 59

background image

59

11 Supplement

VEGAPULS 63 • HART and accumulator pack

40794-EN-130228

Antenna encapsulation Version

Process temperature (measured on the process

fitting)

PTFE

Standard

-40 … +200 °C (-40 … +392 °F)

Low temperature

-200 … +200 °C (-328 … +392 °F)

TFM-PTFE 8 mm

Standard

-40 … +150 °C (-40 … +302 °F)

Low temperature

-196 … +200 °C (-321 … +392 °F)

PFA

Standard

-40 … +200 °C (-40 … +392 °F)

PFA 8 mm

Standard

-40 … +200 °C (-40 … +392 °F)

PTFE with additional pro-

cess seal

FKM

-25 … +130 °C (-13 … +266 °F)

EPDM

-40 … +130 °C (-40 … +266 °F)

Pressure

Version

Process fitting

Vessel pressure

Standard

Flange PN 6

-1 … 6 bar (-100 … 600 kPa/-14.5 … 87 psig)

Flange PN 10

-1 … 10 bar (-100 … 1000 kPa/-14.5 … 145 psig)

Flange PN 16, PN 40

-1 … 16 bar (-100 … 1600 kPa/-14.5 … 232 psig)

Low temperature

Flange DN 50, DN 80,

2", 3"

-1 … 20 bar (-100 … 2000 kPa/-14.5 … 290 psig)

Antenna encapsulation

PFA

Flange ≤ DN 65, 2½"

-1 … 16 bar (-100 … 1600 kPa/-14.5 … 232 psig)

Flange ≥ DN 80, 3"

-0.5 … 16 bar (-50 … 1600 kPa/-7.3 … 232 psig)

Aseptic

SMS

-1 … 6 bar (-100 … 600 kPa/-14.5 … 87 psig)

Tuchenhagen Varivent
Clamp 3", 3½", 4"

-1 … 10 bar (-100 … 1000 kPa/-14.5 … 145 psig)

further hygienic fittings

-1 … 16 bar (-100 … 1600 kPa/-14.5 … 232 psig)

Vibration resistance

4 g at 5 … 200 Hz according to EN 60068-2-6 (vibration

with resonance)

Shock resistance

100 g, 6 ms according to EN 60068-2-27 (mechanical

shock)

Display and adjustment module
Display element

LC display with DOT matrix and backlight

Measured value indication

Ʋ Number of digits

5

Ʋ Size of digits

W x H = 7 x 13 mm

Adjustment elements

4 keys

Protection rating

Ʋ unassembled

IP 20

Ʋ mounted into the housing without

cover

IP 40

Materials

Ʋ Housing

ABS

Ʋ Inspection window

Polyester foil