beautypg.com

VEGA VEGAPULS 63 (≥ 2.0.0 - ≥ 4.0.0) HART and accumulator pack User Manual

Page 58

background image

58

11 Supplement

VEGAPULS 63 • HART and accumulator pack

40794-EN-130228

Gas phase

Temperature

Pressure

1 bar

(14.5 psig)

10 bar

(145 psig)

50 bar

(725 psig)

100 bar

(1450 psig)

200 bar

(2900 psig)

Hydrogen

20 °C/68 °F

-0.01 %

0.10 %

0.61 %

1.2 %

2.5 %

200 °C/392 °F -0.02 %

0.05 %

0.37 %

0.76 %

1.6 %

400 °C/752 °F -0.02 %

0.03 %

0.25 %

0.53 %

1.1 %

Steam (satu-

rated steam)

100 °C/212 °F 0.26 %

-

-

-

-

180 °C/356 °F 0.17 %

2.1 %

-

-

-

264 °C/507 °F 0.12 %

1.44 %

9.2 %

-

-

366 °C/691 °F 0.07 %

1.01 %

5.7 %

13.2 %

76.0 %

Characteristics and performance data
Measuring frequency

K-band (26 GHz technology)

Measuring cycle time

Ʋ Standard electronics approx.

450 ms

Ʋ Electronics with increased sensitivity

approx.

700 ms

Step response time

1)

≤ 3 s

Beam angle

2)

Ʋ Clamp 2", 3"

18°

Ʋ Clamp 3½", 4"

10°

Ʋ Bolting DN 50

18°

Ʋ Bolting DN 80

10°

Ʋ flange DN 50, ANSI 2"

18°

Ʋ Flange DN 80 … DN 150, AN-

SI 3" … 6"

10°

Emitted HF power (depending on the parameter adjustment)

3)

Ʋ Average spectral transmission power

density

-14 dBm/MHz EIRP

Ʋ Max. spectral transmission power

density

+43 dBm/50 MHz EIRP

Ʋ Max. power density in a distance of

1 m

< 1 µW/cm²

Ambient conditions
Ambient, storage and transport tempera-

ture

-40 … +80 °C (-40 … +176 °F)

Process conditions
The following specifications are for information. The specifications on the type plate must be noted.
Temperature

1)

Time span after a sudden measuring distance change by max. 0.5 m in liquid applications, max 2 m with bulk

solids applications, until the output signal has taken for the first time 90 % of the final value (IEC 61298-2).

2)

Outside the specified beam angle, the energy of the radar signal has a level which is reduced by 50 % (-3 dB)

3)

EIRP: Equivalent Isotropic Radiated Power