VEGA Replacement housing for plics instruments User Manual

Page 11

background image

5 Insert the oscillator carefully.

Information:

The electronics of most sensor types is connected via a plug.

Make sure that the plug is in the correct position. The notch

must be in position "18.00 h".

1

Fig. 7: Plug position

1

Notch

6 Screw in and tighten the two screws with a Phillips

screwdriver size 1.

7 Connect the connection cables acc. to the operating

instructions manual of the respective sensor

8 Screw the housing cover back on

The housing exchange is finished.

For Ex applications, the housing exchange must be docu-

mented in general, internally.

Replacement housing for plics instruments

11